Skip to main content

Table 5 Missing screening and its rate of conductivity sensor error at 7 temperature points on SBE-Calibration at JAMSTEC

From: New method of temperature and conductivity sensor calibration with improved efficiency for screening SBE41 CTD on Argo floats

Conductivity (S m−1)

3.0 (1.0 °C)

3.3 (4.5 °C)

4.2 (15.0 °C)

4.6 (18.5 °C)

5.2 (24.0 °C)

5.7 (29.0 °C)

6.0 (32.5 °C)

No. of inconsistency

20

11

5

3

1

3

22

Rate of inconsistency (%)

30.8

16.9

7.7

4.6

1.5

4.6

33.9