Fig. 9From: New method of temperature and conductivity sensor calibration with improved efficiency for screening SBE41 CTD on Argo floatsa Temperature (upper) and salinity (lower) differences (shown in the legend as “dT” and “dS,” respectively) of measurements between SBE41 and SBE3 and 4 sensors during four experiments (Ex1~Ex4) for float S/N 8085 of J-Calibration. J-Calibration system uses cooling packs, adiabatic materials, and a Vinyl greenhouse to keep a uniform environment during the experiments. In 15 min allotted for each measurement, actual calibration is done in 5 min when the dT and dS are within ± 0.01 °C and ± 0.01 PSS-78, respectively. b The same as Fig. 9a except for 5 experiments (Ex1 ~ Ex5) for float S/N F0584 of J-CalibrationBack to article page