From: New method of temperature and conductivity sensor calibration with improved efficiency for screening SBE41 CTD on Argo floats

Fig. 10

Conductivity calibration results to judge whether sensors are acceptable or not. X-axis is conductivity on standard sensor, Y-axis is conductivity differences of SBE41 and standard sensor. and ▲ represent SBE-Calibration and J-Calibration results (S/N 914, 918, and 922), respectively. Colors (dark blue, light blue, and green) indicate sensor S/N (914, 918, and 922), respectively. + means J-Calibration only for APEX (dark orange) and Navis (purple) type floats. The result of SBE calibration for each sensor is shown as a line graph (7 conductivity/temperature points). Area within red dotted lines defines acceptable Argo accuracy, while the area within red dashed lines defines acceptable accuracy requiring re-calibration at least twice