Skip to main content

Table 1 Analytical conditions for major elements and trace F and Cl in olivine by EPMA

From: High precision measurement of trace F and Cl in olivine by electron probe microanalysis

Element

Calibration standard

Accelerating voltage (kV)

Beam current standard (nA)

Beam current sample (nA)

Beam size (jim)

Peak counting time (s)

Background counting time (s)

Analytical crystals

PHA mode (mV)

Bg-(Sinθ  * 105)

(Sinθ *  105)

Si

Rhodonite

20

30

30

5

20

20

TAP

Base line: 1726 Window: 3274

− 1500

1500

Ti

Pure TiO2

20

30

30

5

20

20

LPET

Integral

− 500

500

Al

Pure AI2O3

20

30

30

5

20

20

TAP

Base line: 1347 Window: 3464

− 800

1000

Fe

Pure Fe2O3

20

30

30

5

60

60

LIF

Integral

− 1000

1000

Mg

Pure MgO

20

30

30

5

20

20

TAP

Base line: 1004 Window: 3182

− 2000

2000

Mn

Rhodonite

20

30

30

5

10

10

LIF

Integral

− 800

1000

Ca

Diopside

20

30

30

5

10

10

LPET

Integral

− 600

600

Na

Albite

20

10

30

5

6

6

TAP

Integral

− 800

800

K

K-feldspar

20

10

30

5

10

10

LPET

Integral

− 700

700

F

MgF2

20

10

800

5

240

240

PC1

Base line: 2300 Window: 2170

0

1100

Cl

Halite

20

10

800

5

240 + 240

240 + 240

2 LPET

Integral

− 600

600